Nexus Meter Error Messages
| Message | What It Means | What To Do |
|---|---|---|
| E-b | Appears when the batteries cannot provide enough power for a test. | Replace the batteries immediately. |
| E-U | Appears when a used test strip is inserted. | Repeat with a new test strip. |
| E-E E-0 E-A E-C | Problem in operation. | Repeat the test with a new test strip. If the meter still does not work, please contact GlucoRx Customer Care on 01483 755133 for assistance. |
| E-F | Appears when test strip is removed while counting down, or insufficient blood volume. | Review the instructions an repeat test with a new test strip. If the problem persists, please contact GlucoRx Customer Care on 01483 755133 for help. |
| E-t LOW | Appears when ambient temperature is below system operation range. | System operation range is 10°C to 40°C (50°F to 104°F). Repeat the test after the meter and test strip are in the above temperature range. |
| E-t High | Appears when ambient temperature is above system operation range. | System operation range is 10°C to 40°C (50°F to 104°F). Repeat the test after the meter and test strip are in the above temperature range. |
Troubleshooting
If the meter does not display a message after inserting a test strip...
| Possible Cause | What To Do |
|---|---|
| Batteries exhausted. | Replace the batteries. |
| Test strip inserted upside down or incompletely. | Insert the test strip with contact bar end first and facing up. |
| Defective meter or test strips. | Please contact GlucoRx Customer Care on 01483 755133 |
If the test does not start after applying sample:
| Possible Cause | What To Do |
|---|---|
| Insufficient blood sample. | Repeat the test using a new test strip with larger volume of blood sample. |
| Defective test strip. | Repeat the test with a new test strip. |
| Sample applied after automatic switch-off (3 minutes after last user action). | Repeat the test with a new test strip. Apply sample only when flashing “🌢” appears on the display. |
| Defective meter. | Please contact GlucoRx Customer Care on 01483 755133. |
If the control solution test result is out of range.
| Possible Cause | What To Do |
|---|---|
| Error in performing the test. | Read instructions thoroughly and repeat the test again. |
| Control solution vial was poorly shaken. | Shake the control solution vigorously and repeat the test again. |
| Expired or contaminated control solution. | Check the expiry date of your control solution. |
| Control solution that is too warm or too cold. | Control solution, meter and test strips should be at room temperature (20°C-25°C / 68°F – 77°F) before testing. |
| Defective test strip. | Repeat the test with a new strip. |
| Meter malfunction. | Please contact GlucoRx Customer care on 01483 755133. |
Troubleshooting
Nexus Meter Error Messages
| Message | What It Means | What To Do |
|---|---|---|
| E-b | Appears when the batteries cannot provide enough power for a test. | Replace the batteries immediately. |
| E-U | Appears when a used test strip is inserted. | Repeat with a new test strip. |
| E-E E-0 E-A E-C | Problem in operation. | Repeat the test with a new test strip. If the meter still does not work, please contact GlucoRx Customer Care on 01483 755133 for assistance. |
| E-F | Appears when test strip is removed while counting down, or insufficient blood volume. | Review the instructions an repeat test with a new test strip. If the problem persists, please contact GlucoRx Customer Care on 01483 755133 for help. |
| E-t LOW | Appears when ambient temperature is below system operation range. | System operation range is 10°C to 40°C (50°F to 104°F). Repeat the test after the meter and test strip are in the above temperature range. |
| E-t High | Appears when ambient temperature is above system operation range. | System operation range is 10°C to 40°C (50°F to 104°F). Repeat the test after the meter and test strip are in the above temperature range. |
Troubleshooting
If the meter does not display a message after inserting a test strip...
| Possible Cause | What To Do |
|---|---|
| Batteries exhausted. | Replace the batteries. |
| Test strip inserted upside down or incompletely. | Insert the test strip with contact bar end first and facing up. |
| Defective meter or test strips. | Please contact GlucoRx Customer Care on 01483 755133 |
If the test does not start after applying sample:
| Possible Cause | What To Do |
|---|---|
| Insufficient blood sample. | Repeat the test using a new test strip with larger volume of blood sample. |
| Defective test strip. | Repeat the test with a new test strip. |
| Sample applied after automatic switch-off (3 minutes after last user action). | Repeat the test with a new test strip. Apply sample only when flashing “🌢” appears on the display. |
| Defective meter. | Please contact GlucoRx Customer Care on 01483 755133. |
If the control solution test result is out of range.
| Possible Cause | What To Do |
|---|---|
| Error in performing the test. | Read instructions thoroughly and repeat the test again. |
| Control solution vial was poorly shaken. | Shake the control solution vigorously and repeat the test again. |
| Expired or contaminated control solution. | Check the expiry date of your control solution. |
| Control solution that is too warm or too cold. | Control solution, meter and test strips should be at room temperature (20°C-25°C / 68°F – 77°F) before testing. |
| Defective test strip. | Repeat the test with a new strip. |
| Meter malfunction. | Please contact GlucoRx Customer care on 01483 755133. |
Troubleshooting
Nexus Blue Meter Error Messages
| Message | What It Means | What To Do |
|---|---|---|
| E-b | Appears when the batteries cannot provide enough power for a test. | Replace the batteries immediately. |
| E-U | Appears when a used test strip is inserted. | Repeat with a new test strip. |
| E-E E-0 E-A E-C | Problem in operation. | Repeat the test with a new test strip. If the meter still does not work, please contact GlucoRx Customer Care on 01483 755133 for assistance. |
| E-F | Appears when test strip is removed while counting down, or insufficient blood volume. | Review the instructions an repeat test with a new test strip. If the problem persists, please contact GlucoRx Customer Care on 01483 755133 for help. |
| E-t LOW | Appears when ambient temperature is below system operation range. | System operation range is 10°C to 40°C (50°F to 104°F). Repeat the test after the meter and test strip are in the above temperature range. |
| E-t High | Appears when ambient temperature is above system operation range. | System operation range is 10°C to 40°C (50°F to 104°F). Repeat the test after the meter and test strip are in the above temperature range. |
Troubleshooting
If the meter does not display a message after inserting a test strip...
| Possible Cause | What To Do |
|---|---|
| Batteries exhausted. | Replace the batteries. |
| Test strip inserted upside down or incompletely. | Insert the test strip with contact bar end first and facing up. |
| Defective meter or test strips. | Please contact GlucoRx Customer Care on 01483 755133 |
If the test does not start after applying sample:
| Possible Cause | What To Do |
|---|---|
| Insufficient blood sample. | Repeat the test using a new test strip with larger volume of blood sample. |
| Defective test strip. | Repeat the test with a new test strip. |
| Sample applied after automatic switch-off (3 minutes after last user action). | Repeat the test with a new test strip. Apply sample only when flashing “🌢” appears on the display. |
| Defective meter. | Please contact GlucoRx Customer Care on 01483 755133. |
If the control solution test result is out of range.
| Possible Cause | What To Do |
|---|---|
| Error in performing the test. | Read instructions thoroughly and repeat the test again. |
| Control solution vial was poorly shaken. | Shake the control solution vigorously and repeat the test again. |
| Expired or contaminated control solution. | Check the expiry date of your control solution. |
| Control solution that is too warm or too cold. | Control solution, meter and test strips should be at room temperature (20°C-25°C / 68°F – 77°F) before testing. |
| Defective test strip. | Repeat the test with a new strip. |
| Meter malfunction. | Please contact GlucoRx Customer care on 01483 755133. |
Troubleshooting
Nexus Mini Ultra Meter Error Messages
| Message | What It Means | What To Do |
|---|---|---|
| E-b | Appears when the batteries cannot provide enough power for a test. | Replace the batteries immediately. |
| E-U | Appears when a used test strip is inserted. | Repeat with a new test strip. |
| E-E E-0 E-A E-C | Problem in operation. | Repeat the test with a new test strip. If the meter still does not work, please contact GlucoRx Customer Care on 01483 755133 for assistance. |
| E-F | Appears when test strip is removed while counting down, or insufficient blood volume. | Review the instructions an repeat test with a new test strip. If the problem persists, please contact GlucoRx Customer Care on 01483 755133 for help. |
| E-t LOW | Appears when ambient temperature is below system operation range. | System operation range is 10°C to 40°C (50°F to 104°F). Repeat the test after the meter and test strip are in the above temperature range. |
| E-t High | Appears when ambient temperature is above system operation range. | System operation range is 10°C to 40°C (50°F to 104°F). Repeat the test after the meter and test strip are in the above temperature range. |
Troubleshooting
If the meter does not display a message after inserting a test strip...
| Possible Cause | What To Do |
|---|---|
| Batteries exhausted. | Replace the batteries. |
| Test strip inserted upside down or incompletely. | Insert the test strip with contact bar end first and facing up. |
| Defective meter or test strips. | Please contact GlucoRx Customer Care on 01483 755133 |
If the test does not start after applying sample:
| Possible Cause | What To Do |
|---|---|
| Insufficient blood sample. | Repeat the test using a new test strip with larger volume of blood sample. |
| Defective test strip. | Repeat the test with a new test strip. |
| Sample applied after automatic switch-off (3 minutes after last user action). | Repeat the test with a new test strip. Apply sample only when flashing “🌢” appears on the display. |
| Defective meter. | Please contact GlucoRx Customer Care on 01483 755133. |
If the control solution test result is out of range.
| Possible Cause | What To Do |
|---|---|
| Error in performing the test. | Read instructions thoroughly and repeat the test again. |
| Control solution vial was poorly shaken. | Shake the control solution vigorously and repeat the test again. |
| Expired or contaminated control solution. | Check the expiry date of your control solution. |
| Control solution that is too warm or too cold. | Control solution, meter and test strips should be at room temperature (20°C-25°C / 68°F – 77°F) before testing. |
| Defective test strip. | Repeat the test with a new strip. |
| Meter malfunction. | Please contact GlucoRx Customer care on 01483 755133. |
Troubleshooting
Nexus Voice Meter Error Messages
| Message | What It Means | What To Do |
|---|---|---|
| E-b | Appears when the batteries cannot provide enough power for a test. | Replace the batteries immediately. |
| E-U | Appears when a used test strip is inserted. | Repeat with a new test strip. |
| E-E E-0 E-A E-C | Problem in operation. | Repeat the test with a new test strip. If the meter still does not work, please contact GlucoRx Customer Care on 01483 755133 for assistance. |
| E-F | Appears when test strip is removed while counting down, or insufficient blood volume. | Review the instructions an repeat test with a new test strip. If the problem persists, please contact GlucoRx Customer Care on 01483 755133 for help. |
| E-t LOW | Appears when ambient temperature is below system operation range. | System operation range is 10°C to 40°C (50°F to 104°F). Repeat the test after the meter and test strip are in the above temperature range. |
| E-t High | Appears when ambient temperature is above system operation range. | System operation range is 10°C to 40°C (50°F to 104°F). Repeat the test after the meter and test strip are in the above temperature range. |
Troubleshooting
If the meter does not display a message after inserting a test strip...
| Possible Cause | What To Do |
|---|---|
| Batteries exhausted. | Replace the batteries. |
| Test strip inserted upside down or incompletely. | Insert the test strip with contact bar end first and facing up. |
| Defective meter or test strips. | Please contact GlucoRx Customer Care on 01483 755133 |
If the test does not start after applying sample:
| Possible Cause | What To Do |
|---|---|
| Insufficient blood sample. | Repeat the test using a new test strip with larger volume of blood sample. |
| Defective test strip. | Repeat the test with a new test strip. |
| Sample applied after automatic switch-off (3 minutes after last user action). | Repeat the test with a new test strip. Apply sample only when flashing “🌢” appears on the display. |
| Defective meter. | Please contact GlucoRx Customer Care on 01483 755133. |
If the control solution test result is out of range.
| Possible Cause | What To Do |
|---|---|
| Error in performing the test. | Read instructions thoroughly and repeat the test again. |
| Control solution vial was poorly shaken. | Shake the control solution vigorously and repeat the test again. |
| Expired or contaminated control solution. | Check the expiry date of your control solution. |
| Control solution that is too warm or too cold. | Control solution, meter and test strips should be at room temperature (20°C-25°C / 68°F – 77°F) before testing. |
| Defective test strip. | Repeat the test with a new strip. |
| Meter malfunction. | Please contact GlucoRx Customer care on 01483 755133. |
Troubleshooting
Go Meter Error Messages
| Message | What It Means | What To Do |
|---|---|---|
| E-b | Appears when the batteries cannot provide enough power for a test. | Replace the batteries immediately. |
| E-U | Appears when a used test strip is inserted. | Repeat with a new test strip. |
| E-E E-0 E-A E-C | Problem in operation. | Repeat the test with a new test strip. If the meter still does not work, please contact GlucoRx Customer Care on 01483 755133 for assistance. |
| E-F | Appears when test strip is removed while counting down, or insufficient blood volume. | Review the instructions an repeat test with a new test strip. If the problem persists, please contact GlucoRx Customer Care on 01483 755133 for help. |
| E-t LOW | Appears when ambient temperature is below system operation range. | System operation range is 10°C to 40°C (50°F to 104°F). Repeat the test after the meter and test strip are in the above temperature range. |
| E-t High | Appears when ambient temperature is above system operation range. | System operation range is 10°C to 40°C (50°F to 104°F). Repeat the test after the meter and test strip are in the above temperature range. |
Troubleshooting
If the meter does not display a message after inserting a test strip...
| Possible Cause | What To Do |
|---|---|
| Batteries exhausted. | Replace the batteries. |
| Test strip inserted upside down or incompletely. | Insert the test strip with contact bar end first and facing up. |
| Defective meter or test strips. | Please contact GlucoRx Customer Care on 01483 755133 |
If the test does not start after applying sample:
| Possible Cause | What To Do |
|---|---|
| Insufficient blood sample. | Repeat the test using a new test strip with larger volume of blood sample. |
| Defective test strip. | Repeat the test with a new test strip. |
| Sample applied after automatic switch-off (3 minutes after last user action). | Repeat the test with a new test strip. Apply sample only when flashing “🌢” appears on the display. |
| Defective meter. | Please contact GlucoRx Customer Care on 01483 755133. |
If the control solution test result is out of range.
| Possible Cause | What To Do |
|---|---|
| Error in performing the test. | Read instructions thoroughly and repeat the test again. |
| Control solution vial was poorly shaken. | Shake the control solution vigorously and repeat the test again. |
| Expired or contaminated control solution. | Check the expiry date of your control solution. |
| Control solution that is too warm or too cold. | Control solution, meter and test strips should be at room temperature (20°C-25°C / 68°F – 77°F) before testing. |
| Defective test strip. | Repeat the test with a new strip. |
| Meter malfunction. | Please contact GlucoRx Customer care on 01483 755133. |
Troubleshooting
HCT Meter Error Messages
| Message | What It Means | What To Do |
|---|---|---|
| E-b | Appears when the batteries are too low. | Replace the batteries immediately. |
| E-C | Appears when the wrong code strip is inserted or other coding errors. | Check if the code numbers on the display and ketone foil packet are the same. If the problem persists, please contact GlucoRx Customer Care on 01483 755133 for help. |
| E-2 | Expired code strip. | Repeat with a new test strip. |
| E-U | Appears when a used test strip is inserted. | Repeat with a new test strip. |
| E-t | Appears when ambient temperature is above or below system operation range. | System operation range is 10°C to 40°C(50°F to 104°F). Repeat the test after the meter and test strip are in the above temperature range. |
| E-0 E-A E-E | Problem with the meter. | Repeat the test with a new test strip. If the meter still does not work, please contact GlucoRx Customer Care on 01483 755133 for assistance. |
| E-F | Appears when test strip is removed while counting down, or insufficient blood volume. | Review the instructions and repeat test with a new strip. If the problem persists, please contact GlucoRx Customer Care on 01483 755133 for help. |
Results Readings (for glucose test)
Results Readings (for β-ketone test)
How to interpret blood Ketone results
Reference: adapted from information provided in ‘The Balance Guide to Meds & Kit’, 2011-2012, p15)
| Your ketone reading | Interpretation |
|---|---|
| Below 0.6mmo/L | This is normal |
| Between 0.6 and 1.5mmol/L | You may require assistance; contact your Diabetes healthcare team for advice. |
| Above 1.5mmol/L | Risk of Diabetic ketoacidosis; call your Diabetes healthcare team immediately. |
Troubleshooting
1. If the meter does not display a message after inserting a test strip:
| Possible Cause | What to do |
|---|---|
| Batteries exhausted. | Replace the batteries. |
| Test strip inserted upside down or incompletely. | Insert the test strip with contact bars end first and facing up. |
| Defective meter or test strips. | Please contact GlucoRx Customer Care on 01483 755133. |
2. If the test does not start after applying the sample:
| Possible Cause | What to do |
|---|---|
| Insufficient blood sample. | Repeat the test using a new test strip with larger volume of blood sample. |
| Defective test strip. | Repeat the test with a new test strip. |
| Sample applied after automatic switch-off (3 minutes after last user action). | Repeat the test with a new test strip. Apply sample only when flashing “🌢” appears on the display. |
| Defective meter. | Please contact GlucoRx Customer care on 01483 755133. |
3. If the control testing result is out of range:
| Possible Cause | What to do |
|---|---|
| Error in performing the test. | Read instructions thoroughly and repeat the test again. |
| Control solution vial was poorly shaken. | Shake the control solution vigorously and repeat the test again. |
| Expired or contaminated control solution. | Check the expiry date of the control solution. |
| Control solution that is too warm or too cold. | Control solution, meter, and test strips should be at room temperature 20°C to 25°C (68°F to 77&def;F) before testing. |
| Defective test strip. | Repeat the test with a new test strip. |
| Meter malfunction. | Please contact GlucoRx Customer Care on 01483 755133. |
| Improper working of meter and test strip. | Please contact GlucoRx Customer Care on 01483 755133. |
Q Meter Error Messages
| Message | What It Means | What To Do |
|---|---|---|
| E-b | Appears when the batteries cannot provide enough power for a test. | Replace the batteries immediately. |
| E-U | Appears when a used test strip is inserted. | Repeat with a new test strip. |
| E-E E-0 E-A E-C | Problem in operation. | Repeat the test with a new test strip. If the meter still does not work, please contact GlucoRx Customer Care on 01483 755133 for assistance. |
| E-F | Appears when test strip is removed while counting down, or insufficient blood volume. | Review the instructions an repeat test with a new test strip. If the problem persists, please contact GlucoRx Customer Care on 01483 755133 for help. |
| E-t LOW | Appears when ambient temperature is below system operation range. | System operation range is 10°C to 40°C (50°F to 104°F). Repeat the test after the meter and test strip are in the above temperature range. |
| E-t High | Appears when ambient temperature is above system operation range. | System operation range is 10°C to 40°C (50°F to 104°F). Repeat the test after the meter and test strip are in the above temperature range. |
Troubleshooting
If the meter does not display a message after inserting a test strip...
| Possible Cause | What To Do |
|---|---|
| Batteries exhausted. | Replace the batteries. |
| Test strip inserted upside down or incompletely. | Insert the test strip with contact bar end first and facing up. |
| Defective meter or test strips. | Please contact GlucoRx Customer Care on 01483 755133 |
If the test does not start after applying sample:
| Possible Cause | What To Do |
|---|---|
| Insufficient blood sample. | Repeat the test using a new test strip with larger volume of blood sample. |
| Defective test strip. | Repeat the test with a new test strip. |
| Sample applied after automatic switch-off (3 minutes after last user action). | Repeat the test with a new test strip. Apply sample only when flashing “🌢” appears on the display. |
| Defective meter. | Please contact GlucoRx Customer Care on 01483 755133. |
If the control solution test result is out of range.
| Possible Cause | What To Do |
|---|---|
| Error in performing the test. | Read instructions thoroughly and repeat the test again. |
| Control solution vial was poorly shaken. | Shake the control solution vigorously and repeat the test again. |
| Expired or contaminated control solution. | Check the expiry date of your control solution. |
| Control solution that is too warm or too cold. | Control solution, meter and test strips should be at room temperature (20°C-25°C / 68°F – 77°F) before testing. |
| Defective test strip. | Repeat the test with a new strip. |
| Meter malfunction. | Please contact GlucoRx Customer care on 01483 755133. |